*CS SERIES Control Systems are used on all Taylor-Wharton LABS units.
The unit featured is only one of the cryostorage systems available in the CryoScience by Taylor-Wharton™ lines.
CS SERIES
CS200 OPERATION & MAINTENANCE INSTRUCTIONS
Control System
REVIEW AND UNDERSTAND ALL SAFETY PROCEDURES IN THE “HANDLE WITH CARE”
BOOKLET FORM # TW-10 P/N 7950-8052 BEFORE ATTEMPTING TO INSTALL, OPERATE
OR PERFORM MAINTENANCE ON THIS LN
2
CONTROLLER.
DO NOT ATTEMPT TO USE OR MAINTAIN ANY LIQUID NITROGEN FREEZER UNTIL YOU
READ AND UNDERSTAND THESE INSTRUCTIONS.
DO NOT PERMIT UNTRAINED PERSONS TO USE OR MAINTAIN THIS UNIT.
IF YOU DO NOT FULLY UNDERSTAND THESE INSTRUCTIONS, CONTACT YOUR SUPPLIER
FOR FURTHER INFORMATION.
BEFORE ATTEMPTING TO OPERATE THIS CONTROLLER WITH ANY CRYOSCIENCE BY
TAYLOR-WHARTON™ LABS & K SERIES UNITS, YOU MUST READ THE SEPARATE
OPERATING AND SAFETY MANUAL PROVIDED WITH THAT TAYLOR-WHARTON UNIT.
CAUTION - SAFETY FIRST!
Page view 0
1 2 3 4 5 6 ... 53 54

Summary of Contents

Page 1 - CS SERIES

*CS SERIES Control Systems are used on all Taylor-Wharton LABS units. The unit featured is only one of the cryostorage systems available in the CryoSc

Page 2

8CS SERIESIf possible, the original packing material should be retained for reshipment. If not available, consult Taylor-Wharton for shipping and pack

Page 3 - TABLE OF CONTENTS

9CS SERIESEN Compliance tables found on pages 43-46. Care should be taken to use only manufacturer-recommended cables, power supplies and electrical a

Page 4 - (continued...)

10CS SERIESControl Components CS SERIES Control System ComponentsThe CS SERIES Control System for the Taylor-Wharton Cryostorage freezer consists of t

Page 5 - SAFETY PRECAUTIONS

11CS SERIESOperational TheoryThe CS SERIES Control System automatically maintains the Liquid Nitrogen (LN2) level and monitors temperature in the Cryo

Page 6 - GENERAL INFORMATION

12CS SERIESThe LN2 level is indicated on the display and is determined by the number of thermistors submerged in LN2 according to the table below:Tabl

Page 7

13CS SERIESTable 2.0 Alarm ConditionsALARM PROBLEM DETECTEDOperating in Battery ModeAlarm WarningThe power from the power supply has been disrupted an

Page 8

14CS SERIESPower Supply A 12 VDC power supply is supplied for the CS SERIES Control System. The system is supplied with a transformer compatible with

Page 9 - DELIVERY AND RETURNS

15CS SERIESTemperature MonitoringThe CS SERIES Control System uses a type T thermocouple to monitor the temperature in the vessel. The thermocouple is

Page 10 - INSTALLATION

16CS SERIESIn addition, inspection and preventive maintenance should also be performed on the freezer and its mechanical parts. Refer to owner’s manua

Page 11 - OPERATION

17CS SERIESControl Setting AdjustmentsLevel The standard sensor assembly that is installed on a freezer consists of a circuit board with thermistors i

Page 13 - Operational Theory

18CS SERIESView Only The View Only menu selection allows the user to view settings but will not allow changes to any settings. This restriction is des

Page 14 - ALARM PROBLEM DETECTED

19CS SERIES – Sensor Type: Allows the user to set the sensor type as a 4 thermistor sensor or an 8 thermistor sensor. Also allows the user to disable

Page 15

20CS SERIES•Remote Alarm Delay: Allows the user to set the time before the remote alarm relay is triggered after an error condition occurs. Settings

Page 16 - Operating Parameters

21CS SERIESTemperature Control The CS SERIES Control System has the capability to control the vapor temperature at the top of the freezer. The tempera

Page 17 - Liquid Phase Storage

22CS SERIESMENUVIEW ONLYTEMPERATURELEVELSYSTEMSCHEDULED EVENTSTHERMOCOUPLE SELECTTEST TEMPERATURE SYSTEMCONTROL BY TEMPERATURETEMPERATURE ALARMSTEMPER

Page 18 - CONTROLLER OPERATION

23CS SERIESCHANGE SETTINGSCONTACT SUPPORTSYSTEM INFORMATIONTEMPERATURELEVELSYSTEMSCHEDULED EVENTSSECURITYDISTRIBUTORPAGER DIGITALTAYLOR-WHARTONSERIAL

Page 19 - Control Setting Adjustments

24CS SERIESFactory DefaultsThermocouple #1 OnThermocouple #2 OffControl by Temperature OffHigh Temperature Alarm #1 -100CHigh Temperature Alarm #2

Page 20

25CS SERIESTroubleshootingIf the Taylor-Wharton Cryostorage freezer with the CS200 SERIES Control System installed experiences problems or appears tha

Page 21

26CS SERIESWiring Diagram4 wires position 1 P1P14 wires position 2 P1P210 wires position 1 P2P110 wires position 2 P2P2Blue P1P1 ValveRed P1P1Yellow +

Page 22 - Temperature

27CS SERIESInstallation & SetupThe CS SERIES Control System consists of the following components.• Main Control• VFD Display Module • Wiring Ha

Page 23 - Lid Switch

CS SERIESTABLE OF CONTENTSSAFETY PRECAUTIONS ... 3Liquid Nitrogen ...

Page 24

28CS SERIESConnect the level sensor assembly into the pigtail with the round connector terminating at the control end of the harness assembly. The con

Page 25

29CS SERIESConnect the Remote Alarm plug into the panel at the end of the wiring harness assembly. This is keyed so that it can only be plugged in one

Page 26 - Factory Defaults

30CS SERIESTemperature Thermocouple Select The chamber temperature is monitored with 1 or 2 Type T thermocouples. The thermocouple is placed in the ch

Page 27 - Interconnection Block Diagram

31CS SERIESTable 4.0 Temperature Calibration InstructionAccess the menu by pressingHighlight CHANGE SETTINGS and pressHighlight TEMPERATURE and press

Page 28 - Wiring Diagram

32CS SERIESAlarms and Error Conditions The CS SERIES Control System control tracks many different conditions in the freezer and therefore, has a full

Page 29 - Connection to Main Control

33CS SERIESLoggingOn board memory logging function is one of the most powerful and useful features of the CS SERIES Control System control. It provide

Page 30

34CS SERIESDUMP ERROR LOGS is accessible through the menu system of the control. This option sends data from the error logs to the serial port of the

Page 31 - External Connector Ratings

35CS SERIESErase LogsERASE LOGS is accessible through the menu system of the control. This menu choice erases any of the four logs found in the contro

Page 32 - Temperature Calibration

36CS SERIESRemoving/Installing the Solenoid Valve The LN2 and power must be turned off before beginning work on the solenoid valve.For LABS UnitsDisco

Page 33 - Test Level Sensors

37CS SERIESPlumbing AssemblySolenoid Valve90 deg Elbow90 deg ElbowRelief Valve TubeFemale ConnectorRelief ValveStrainerReducerMale BranchTeeMale Branc

Page 34 - Test Alarms

CS SERIESError Logs ...34System Logs ...

Page 35 - Dump Logs

38CS SERIESIndicates High Liquid Level 1. Determine actual liquid level using a dipstick. Select MENU, LEVEL SENSING, TEST LEVEL SENSORS. An “L” or “

Page 36 - Temperature Logs

39CS SERIESIndicates Low LN2 SupplyWhen the START FILL thermistor is uncovered, the controller calls for the fill solenoid valve to open. If the STOP F

Page 37 - Assembly

40CS SERIESFill Solenoid Cycles On and OffCS SERIES Control System features a timer function whereby the valve open duration is limited, allowed to sh

Page 38 - Controller Electrical Tests

41CS SERIESQCF (Quick Chill Feature) Will Not Operate See LID OPEN ALARM To activate or adjust times: Select MENU, USER OPTIONS, CONTROL OPTIONS, LID/

Page 39 - TROUBLESHOOTING

42CS SERIESSERVICE AND MAINTENANCE HISTORY LOGFill in top section at installation. Copy this form each time service is required. Fill bottom section

Page 40 - Indicates High Liquid Level

43CS SERIESAppedixCertifications & ListingsThis product complies with the following standards and directives: • CB Scheme Report and Certificate •

Page 41 - Indicates Open Sensor

44CS SERIESTable 7.0Guidance and Manufacturer’s Declaration – Electromagnetic ImmunityThe CS CONTROL SYSTEM is intended for use in the electromagnetic

Page 42 - Lid Open Alarm

45CS SERIESTable 8.0Guidance and Manufacturer’s Declaration – Electromagnetic ImmunityThe CS CONTROL SYSTEM is intended for use in the electromagnetic

Page 43 - REPLACEMENT PARTS

46CS SERIESTable 9.0Recommended separation distances between portable and mobile RF communications equipment and the CS CONTROL SYSTEM.The CS CONTROL

Page 44

47CS SERIESDeclaration of ConformityEC DECLARATION OF CONFORMITYThe undersigned representing the manufacturer Pacer Digital Systems, Inc. Attn: Kevin

Page 45 - EN Compliance Tables

3CS SERIESSAFETY PRECAUTIONSLiquid Nitrogen Nitrogen is an inert, colorless, odorless, and tasteless gas making up four-fifths of the air you breathe –

Page 46 - Table 7.0

48CS SERIESThis warranty is also void if the product is in any way altered or repaired by others. Taylor-Wharton shall not be liable under this warran

Page 47 - Table 8.0

49CS SERIESCryowireNetworkBattery12V, 2.5ACryowireNetworkNOCNCCryowireLocalPower12 VDC, 3ARemoteAlarm4075 Hamilton Blvd.Theodore, AL 365821-800-898-26

Page 50

CS Series Control System – 05/13 – P/N 7950-8361 ©2013 Taylor-Wharton International. All rights reserved.TW Cryogenics LLC, a wholly owned subsidiary

Page 51

4CS SERIESFor more detailed information concerning safety precautions and safe practices to be observed when handling cryogenic liquids consult CGA pu

Page 52

5CS SERIESCS SERIES Control System SpecificationsControl Type: LN2 Level Control & Temperature Control Level Measurement Sensor Type: 8-Thermisto

Page 53

6CS SERIESAlarms (continued...): Low Temperature Alarm: (T/C #1, T/C#2) Programmable Thermocouple Calibration: Alarm Always enabled Thermocouple Op

Page 54 - Australia and Asia

7CS SERIESDELIVERY AND RETURNSUnpacking and InspectionInspect shipping containers for external damage. All claims for damage (apparent or concealed) o

Comments to this Manuals

No comments